液液萃取结合气相色谱–电子捕获检测法(LLE-GC/ECD)测定集中式生活饮用水源地地表水中7种多氯联苯
Determination of Seven Polychlorinated Biphenyls in Surface Water of Centralized Drinking Water Sources by Liquid-Liquid Extraction Coupled with Gas Chromatography-Electron Capture Detection (LLE-GC/ECD)
摘要: 多氯联苯(PCBs),是一类有机化合物,集中式生活饮用水源地地表水中多氯联苯包含PCB1016、PCB1221、PCB1232、PCB1242、PCB1248、PCB1254、PCB1260。本文采用液液萃取结合气相色谱–电子捕获检测法(LLE-GC/ECD)测定集中式生活饮用水源地地表水中多氯联苯的含量。结果表明:该方法测定集中式生活饮用水源地地表水中多氯联苯,有很好的校正曲线(R > 0.995),取样量为1.0 L时,检出限为0.0008~0.0058 μg/L,平均加标回收率为99.5%~120.0%,相对标准偏差(n = 9)为0.6%~4.3%。本方法具有操作简便、灵敏度高和重复性好等特点,为环境监测部门开展饮用水源地多氯联苯的日常监测提供了可靠的分析方法参考。
Abstract: Polychlorinated biphenyls (PCBs) are a class of organic compounds. The PCBs in surface water of centralized drinking water sources include PCB1016, PCB1221, PCB1232, PCB1242, PCB1248, PCB1254, and PCB1260. This paper uses liquid-liquid extraction combined with gas chromatography-electron capture detection (LLE-GC/ECD) to determine the content of PCBs in surface water of centralized drinking water sources. The results show that this method has a good calibration curve (R > 0.995) for the determination of PCBs in surface water of centralized drinking water sources. When the sample volume is 1.0 L, the detection limit is 0.0008~0.0058 μg/L, the average spiked recovery rate is 99.5%~120.0%, and the relative standard deviation (n = 9) is 0.6%~4.3%. This method is characterized by simplicity, high sensitivity, and good repeatability, providing a reliable analytical method reference for environmental monitoring departments to conduct routine monitoring of PCBs in drinking water sources.
文章引用:张兰普, 邵菠昌. 液液萃取结合气相色谱–电子捕获检测法(LLE-GC/ECD)测定集中式生活饮用水源地地表水中7种多氯联苯[J]. 环境保护前沿, 2025, 15(3): 369-377. https://doi.org/10.12677/aep.2025.153044

1. 引言

多氯联苯(PCBs),是一类有机化合物,按氯原子数或氯的百分含量分别加以标号。纯化合物为结晶态,混合物则为油状液体。低氯化物呈液态,流动性好,随着氯原子数的增加,粘稠度相应增高,呈糖浆状乃至树脂状[1]。多氯联苯是由209种化学物质组成的家族,具有半挥发性,能够从水体或土壤中以蒸气形式进入大气环境或被大气颗粒物吸附,通过大气环流远距离迁移[2]。具有较好的稳定性、耐热性以及绝缘性,可用作润滑材料、增塑剂、杀菌剂、热载体和变压器油等,被广泛地应用在电器的绝缘油、感压纸等各种领域。世界卫生组织国际癌症研究机构公布的致癌物清单参考,多氯联苯在1类致癌物清单中[3]。对皮肤、牙齿、神经行为、免疫功能、肝脏有影响,且具有生殖毒性和致畸性、致癌性[4]。多氯联苯在GB 3838-2002《地表水环境质量标准》中为集中式生活饮用水源地地表水源地特定项目,包含PCB1016、PCB1221、PCB1232、PCB1242、PCB1248、PCB1254、PCB1260七种,规定限值为0.02 μg/L [5]。目前,测定水体多氯联苯的方法中,GB 3838-2002《地表水环境质量标准》推荐的方法《水和废水标准检验法(第15版)》[6]中方法操作繁琐,没有相应的检出限,对实际监测工作参考价值不高,这也给日常监测工作带来不便。本文采用液液萃取结合气相色谱–电子捕获检测法(LLE-GC/ECD)测定集中式生活饮用水源地地表水中7种多氯联苯的含量,该方法具有操作简便、灵敏度高和重复性好等特点,满足集中式生活饮用水源地地表水中7种多氯联苯的测定,为环境监测部门开展饮用水源地多氯联苯的日常监测提供了可靠的分析方法参考。

2. 实验环节

2.1. 仪器、耗材和试剂

氮吹浓缩仪;分液漏斗振荡器;Agilent 7890A气相色谱仪,具分流/不分流进样口,配制电子捕获检测器(ECD);Agilent 19091S-433 HP-5MS 30 m × 0.25 mm × 0.25 μm气相色谱柱;微量注射器:5、10、25、50、100、250、500 µl和1 ml;分液漏斗:2 L;全自动固相萃取装置及一般实验室常用仪器和设备。

氮气,纯度 ≥ 99.999%;PCBs标准储备液:100 mg/L (正己烷溶剂),Accustandard,Inc.将上述储备液用正己烷稀释为质量浓度为5 mg/L PCBs标准使用溶液;正己烷(色谱纯);丙酮(农残级);氯化钠与无水硫酸钠:均为优级纯,使用前用马弗炉于400℃烘烤6 h,冷却后保存于洁净的试剂瓶,将试剂瓶放于干燥皿中;硅酸镁固相小柱。

2.2. 仪器分析条件

本研究采用配备电子捕获检测器(ECD)的Agilent 7890A气相色谱系统进行分析。色谱分离在Agilent HP-5MS毛细管色谱柱(30 m × 0.25 mm × 0.25 μm,部件号:19091S-433)上完成,载气选用高纯氮气(99.999%),流速恒定于1.0 mL/min。进样系统设置为不分流模式,进样体积1.0 μL,进样口与检测器温度均维持在280℃。程序升温条件设定如下:初始温度100℃ (无保持时间),以8℃/min升温至180℃后立即以5℃/min升至220℃,最终温度保持20 min (针对PCB1260同系物分析时保持时间延长至60 min)。实验前按标准操作流程启动系统:首先开启高纯氮气气源,确认压力稳定在0.6 MPa后启动色谱主机。系统完成初始化自检后执行基线稳定程序,持续运行2~3小时至检测信号基线漂移值稳定在300 μV以下,确认系统性能参数符合分析方法验证要求后方可进行样品分析。

2.3. 样品

2.3.1. 样品的采集

采样瓶为具聚四氟乙烯衬里的1L棕色玻璃瓶,采样瓶使用前应用洗涤剂洗涤,然后用水冲洗,再用丙酮清洗一次,最后用二氯甲烷或正己烷清洗一次,400℃烘干。采样时用采样瓶直接采集样品至瓶满为止,不要用样品洗涤采样瓶。每批样品应带一个全程序空白,每10个样品需采集一个平行样。采集完水样后应在样品瓶上立即贴上标签,4℃保存。如果水中有余氯,则每升样品中加入80 mg硫代硫酸钠。

2.3.2. 样品的保存

样品在4℃冰箱中可保存7天。萃取后如暂不分析,样品可保存40天(<−20℃条件下避光存放)。

2.3.3. 试样的制备–液液萃取

试样前处理方式采用液液萃取,其原理是利用物质在两个不混溶的溶剂中溶解度不同来进行分离,影响方式包括搅拌或震荡强度和接触时间,通过反复多次实验,本实验萃取条件为,量取1 L水样倒入2 L的分液漏斗,加入30 g NaCl,轻轻振摇直至NaCl完全溶解。往分液漏斗中加入60 ml正己烷,振幅40 mm,转速为260转/min,垂直震荡萃取5 min,时间过短可能导致萃取不完全,而时间过长则可能引入其他杂质或导致溶剂损失,并注意放气。静置10 min,如静置时间过短,会导致溶剂损失,导致萃取不完全,使有机相分层。将正己烷放出,使用三角烧瓶收集。为提高萃取效率,再往水相中加入60 ml正己烷,重复2次上述液液萃取过程。合并3次正己烷萃取液。

将有机相通过适量无水硫酸钠脱水,收集于浓缩瓶中,氮吹浓缩至0.5 ml左右,转移到进样小瓶中,浓缩瓶用少量正己烷洗涤,洗涤液转移到进样小瓶中,用正己烷定容至1 ml,进行GC-ECD分析。

如果样品严重乳化,需要采用如搅动、离心、冷冻、玻璃棉过滤等手段完成两相分离。可根据样品乳化程度选择适合的破乳方法。对于轻度乳化的样品,可以在水平方向上用玻璃棒搅动或缓慢地旋转摇动分液漏斗,消除界面处的“泡沫”,促进分层。对于中重度乳化的样品,可将乳化混合物移入离心分离机中,进行高速离心分离。实验证明该方法效果明显且省时;也可以将乳浊液加盖放置冷冻,乳浊液会分离成有机层和冰冻层两层。这种方法虽然有效,但对于赶进度的实验可能不太适用。还可以使用玻璃棉过滤或直接过滤,以去除乳化层。若水样中的悬浮物>1%,需要静置分离,分离后取上清液按上述方法进行提取。

2.4. 标准曲线的绘制

质量浓度为5 mg/L PCBs (PCB1016、PCB1221、PCB1232、PCB1242、PCB1248、PCB1254、PCB1260)标准使用溶液。分别取上述标准使用溶液10 μL、20 μL、40 μL、80 μL、100 μL用正己烷定容到1.0 mL,配制得到浓度为50 μg/L、100 μg/L、200 μg/L、400 μg/L、500 μg/L标准系列质量浓度。

按照1.2仪器分析条件,将配制好的5个浓度校准曲线系列样品由低浓度到高浓度依次测定,在每种PCBs指纹峰中选择3个定量峰分别进行定量,取这3个定量峰定量结果的平均值为最终定量结果。以目标物的质量浓度为横坐标,峰面积为纵坐标,建立标准曲线。

3. 结果与讨论

3.1. 多氯联苯色谱图

在本方法规定的条件下,目标化合物的色谱图见图1~7。由图1~7可知,7种多氯联苯峰尖锐且分离良好。

3.2. 标准曲线方程、检出限及测定下限

按照《环境监测分析方法标准制订技术导则》(HJ 168-2020)附录A.1规定,进行方法检出限和测定下限分析。

Figure 1. Chromatogram of the standard sample of PCB1016

1. PCB1016标准样品色谱图

Figure 2. Chromatogram of the standard sample of PCB1221

2. PCB1221标准样品色谱图

Figure 3. Chromatogram of the standard sample of PCB1232

3. PCB1232标准样品色谱图

Figure 4. Chromatogram of the standard sample of PCB1242

4. PCB1242标准样品色谱图

Figure 5. Chromatogram of the standard sample of PCB1248

5. PCB1248标准样品色谱图

Figure 6. Chromatogram of the standard sample of PCB1254

6. PCB1254标准样品色谱图

Figure 7. Chromatogram of the standard sample of PCB1260

7. PCB1260标准样品色谱图

空白试验中均未检出7种PCBs,按照标准方法HJ 168要求,对浓度含量为0.040 μg/L的1 L空白加标样品进行11次重复性测定。将测定结果,按公式(1)计算方法检出限。以4倍的检出限作为测定下限,即

R Q L = 4 × M D L M D L = t ( n 1 , 0.99 ) × S (1)

式中:MDL——方法检出限;t (10,0.99)——自由度为 n 1 (n为测定次数11),置信度为99%时的t分布(单侧)的系数,值为2.764;S-11次重复性测定的标准偏差。11次测定结果见表1。由表1可知,7种多氯联苯具有良好的线性,相关系数r ≥ 0.995,检出限在0.0008~0.0058 μg/L之间。

Table 1. Results of method standard curve equation, detection limit and determination lower limit

1. 方法标准曲线方程、检出限及测定下限结果

项目

校准曲线方程

相关系数(r)

0.040 μg/L

方法检出限(μg/L)

测定下限(μg/L)

平均值

标准偏差S

PCB1016

y = 1.03269x + 10.86574

0.9992

0.039

0.0010

0.0028

0.0112

y = 1.34449x + 12.14087

0.9994

y = 1.00195x + 6.10196

0.9995

PCB1221

y = 0.496055x + 4.90539

0.9994

0.038

0.0013

0.0036

0.0144

y = 0.338543x + 2.74871

0.9994

y = 1.13306x + 10.86574

0.9988

PCB1232

y = 0.915728x + 1021658

0.9988

0.040

0.0021

0.0058

0.0232

y = 0.497336x + 4.64862

0.9989

y = 0.608055x + 6.64436

0.9987

PCB1242

y = 0.797285x + 8.72246

0.9989

0.039

0.0003

0.0008

0.0032

y = 0.998154x + 11.48937

0.9990

y = 1.05750x + 8.48657

0.9993

PCB1248

y = 1.17428x + 17.13456

0.9988

0.041

0.0006

0.0017

0.0068

y = 1.41409x +17.26796

0.9991

y = 1.72597x + 21.10192

0.9991

PCB1254

y = 1.50434x +22.53085

0.9981

0.038

0.0010

0.0028

0.0112

y = 2.16001x +34.65372

0.9981

y = 2.52206x +28.77701

0.9989

PCB1260

y = 2.97014x +47.05863

0.9983

0.038

0.0005

0.0014

0.0056

y =3.28552x +57.90560

0.9980

y = 4.24441x + 72.17667

0.9981

3.3. 精密度、正确度分析

对7种PCBs浓度含量为0.060 μg/L、0.160 μg/L、0.200 μg/L、0.400 μg/L的1 L空白加标样品进行9次重复性测定。计算相对标准偏差与加标回收率,测定结果见表2。由表2可知,对7种PCBs准确度在98.5%~110%之间,精密度在0.6%~4.3%之间;具有良好的准确度和精密度。

Table 2. Results of precision and accuracy analysis

2. 精密度、正确度分析结果

项目名称

0.060 μg/L

0.160 μg/L

0.200 μg/L

0.400 μg/L

均值

RSD%

平均 加标

回收率%

均值

RSD%

平均 加标

回收率%

均值

RSD%

平均 加标

回收率%

均值

RSD%

平均 加标

回收率%

PCB1016

0.064

1.3

107

0.161

1.5

101

0.196

1.7

98.5

0.401

1.9

100

PCB1221

0.061

2.1

102

0.160

0.9

100

0.199

2.4

99.5

0.400

1.4

100

PCB1232

0.060

3.1

100

0.162

2.4

101

0.206

3.6

103

0.410

1.0

102

PCB1242

0.064

1.0

107

0.167

2.3

104

0.212

3.3

106

0.411

1.1

103

PCB1248

0.066

1.5

110

0.166

0.6

104

0.208

1.2

104

0.418

2.6

104

PCB1254

0.063

4.3

105

0.162

1.3

101

0.212

2.1

106

0.408

1.7

102

PCB1260

0.062

1.8

103

0.165

2.7

103

0.201

2.1

100

0.399

1.1

99.8

4. 实际样品测定

按照本方法的要求,采集我市某集中式生活饮用水源地地表水源地的实际样品按照1.3步骤进行测定。实际样品测定结果见表3

Table 3. Results of actual sample determination

3. 实际样品测定结果

监测项目

样品类型

样品测定结果 (μg/L)

平行样测定结果(μg/L)

加标样(0.200 μg/L) 测定结果(μg/L)

加标回收率(%)

PCB1016

集中式生活

饮用水地表水

未检出

未检出

0.217

108

PCB1221

未检出

未检出

0.204

102

PCB1232

未检出

未检出

0.241

120

PCB1242

未检出

未检出

0.199

99.5

PCB1248

未检出

未检出

0.221

112

PCB1254

未检出

未检出

0.221

110

PCB1260

未检出

未检出

0.209

104

表3可见在所测试的集中式生活饮用水源地地表水中均未检出多氯联苯。回收率在99.5%~120%之间;有较高的准确度,具有很强的实用性,符合分析测试要求,适用于集中式生活饮用水源地地表水中7种多氯联苯的测定。

5. 结语

本研究建立了液液萃取结合气相色谱–电子捕获检测法(LLE-GC/ECD)测定集中式生活饮用水源地地表水中7种多氯联苯的分析方法。通过系统优化实验条件,开展方法验证与实际水样检测,实验所需仪器、标准物质及关键试剂均为实验室常规配置或市售产品。严格实施的质量控制措施有效保障了数据的准确性,方法验证结果显示:方法检出限为0.0008~0.0058 μg/L,测定下限0.0032~0.0232 μg/L,加标回收率99.5%~120%,相对标准偏差 < 5%,本方法具有操作简便、灵敏度高和重复性好等特点,为环境监测部门开展饮用水源地多氯联苯的日常监测提供了可靠的分析方法参考。

NOTES

*第一作者。

#通讯作者。

参考文献

[1] 王箴. 化工辞典[M]. 北京: 化学工业出版社, 2010: 186.
[2] 环境保护部. 国家污染物环境健康风险名录——化学第一分册[M]. 北京: 中国环境科学出版社, 2009.
[3] International Agency for Research on Cancer (2018) IARC Monographs Volume 118: Polychlorinated Biphenyls and Polybrominated Biphenyls. IARC Press.
https://publications.iarc.fr/Book-And-Report-Series/Iarc-Monographs-On-The-Identification-Of-Carcinogenic-Hazards-To-Humans/Polychlorinated-Biphenyls-And-Polybrominated-Biphenyls-2018
[4] 丁晓雯, 柳春红. 食品安全学[M]. 北京: 中国农业大学出版社, 2011: 140-145.
[5] 国家环境保护总局, 国家质量监督检验检疫总局. 地表水环境质量标准: GB3838-2002 [S]. 北京: 中国环境出版集团出版社, 2021.
[6] 水和废水标准检验法[M]. 第15版. 北京: 中国建筑工业出版社, 1985.