本文针对温湿度标准箱温度均匀性对湿度校准结果的影响,提出一种通过露点温度修正相对湿度的方法,采用精密数字温湿度计、湿度发生器验证该方法的合理性,验证且提升湿度校准的工作的准确性。 This paper studies the influence of temperature uniformity of temperature and humidity standard box on humidity calibration results, and proposes a method to correct relative humidity through dew point temperature. It uses precision digital hygrograph and humidity generator to verify the rationality of the method and verify and improve the accuracy of humidity calibration.
本文针对温湿度标准箱温度均匀性对湿度校准结果的影响,提出一种通过露点温度修正相对湿度的方法,采用精密数字温湿度计、湿度发生器验证该方法的合理性,验证且提升湿度校准的工作的准确性。
温度均匀性,相对湿度,露点
Yuming Liu, Xin Wu
Shanghai Institute of Metrology and Testing Technology, Shanghai
Received: Feb. 19th, 2024; accepted: Feb. 29th, 2024; published: Mar. 15th, 2024
This paper studies the influence of temperature uniformity of temperature and humidity standard box on humidity calibration results, and proposes a method to correct relative humidity through dew point temperature. It uses precision digital hygrograph and humidity generator to verify the rationality of the method and verify and improve the accuracy of humidity calibration.
Keywords:Temperature Uniformity, Relative Humidity, Dew Point
Copyright © 2024 by author(s) and beplay安卓登录
This work is licensed under the Creative Commons Attribution International License (CC BY 4.0).
http://creativecommons.org/licenses/by/4.0/
温湿度标准箱为数字式温湿度计的校准工作提供指定的温度、湿度环境。《JJF1076-2020数字式温湿度计校准规范》 [
在日常工作中,部分温湿度标准箱的温度均匀性随相对湿度的减小而增大。该标准箱使用《JJF1564-2016温湿度标准箱校准规范》进行校准时,符合《JJF1076-2020数字式温湿度计校准规范》的技术要求。但实际在低湿工况(例如10% RH)进行校准工作时,均匀性远大于技术要求,温度偏差引入湿度的校准误差。
校准工作中,精密露点仪 [
本文选用相对湿度精度为2% RH数字温湿度计作为被校准仪器,20℃条件下20% RH、30% RH、40% RH、50% RH、60% RH、70% RH、80% RH和90% RH作为试验工况。将数字温湿度计放置在温湿度标准箱内,该温度均匀性大、温湿度计距离标准设备较远,分别依次设定上述试验工况,待温湿度标准箱和数字温湿度计达到稳定后30 min,每隔2 min记录温度和相对湿度标准值、仪器温度和相对湿度示值,每个工况记录三组数据并取平均值。
温度、露点温度和相对湿度三者呈函数关系 [
湿度发生器的校准室体积小、温湿度均匀性极小。将被校准仪器放置在湿度发生器内,依次设定上述工况,达到稳定后30 min,每隔2 min记录的校准数据作为标准值。对比被校准仪器在温湿度标准箱中测得值、修正后的换算值、湿度发生器中的测得值三个湿度数据,分析该修正方法的合理性和准确性。
不同温度工况点,相对湿度与温度的试验数据如表1、表2、表3所示。
湿度点 | T标 | T仪器 | Td | % RH标 | % RH示 | % RH换算 |
---|---|---|---|---|---|---|
20% RH | 19.91 | 20.3 | −4.0 | 18.82 | 17.9 | 19.26 |
30% RH | 20.12 | 20.5 | 1.6 | 29.12 | 28.0 | 28.70 |
40% RH | 20.12 | 20.4 | 5.8 | 39.16 | 38.5 | 39.17 |
50% RH | 20.18 | 20.3 | 9.2 | 49.22 | 48.5 | 49.16 |
60% RH | 20.10 | 20.2 | 12.2 | 60.40 | 59.7 | 60.44 |
70% RH | 20.05 | 20.2 | 14.6 | 70.85 | 70.2 | 70.85 |
80% RH | 20.03 | 20.2 | 16.7 | 81.15 | 80.1 | 80.95 |
90% RH | 20.04 | 20.2 | 18.4 | 90.29 | 89.3 | 90.19 |
表1. 温湿度标准箱测试数据统计
湿度点 | ΔT (T仪器-T标) | Δ%RH1(|% RH仪器-%RH标|) | Δ% RH2(|% RH换算-% RH标|) |
---|---|---|---|
20% RH | 0.39 | 0.9 | 0.4 |
30% RH | 0.38 | 1.1 | 0.4 |
40% RH | 0.28 | 0.7 | 0.0 |
50% RH | 0.12 | 0.7 | 0.1 |
60% RH | 0.10 | 0.7 | 0.0 |
70% RH | 0.15 | 0.6 | 0.0 |
80% RH | 0.17 | 1.1 | 0.2 |
90% RH | 0.16 | 1.0 | 0.1 |
表2. 无线温湿度数据采集仪温度、相对湿度差值统计
表1统计了温湿度标准箱测试的标准器温度与相对湿度和采集仪显示的温度与相对湿度,以及未修正和修正相对湿度。温度与两种相对湿度与标准器标准值的差值如表2所示,图1为差值折线图。
结合表1和表2可以发现,湿度越高箱体实际温度越接近;湿度低于60% RH的温度偏移20℃超过0.1℃,且随着湿度降低采集仪的实际温度与标准器实际温度偏移温度ΔT越来越大,20% RH时达0.39摄氏度。未修正相对湿度的Δ% RH1基本高于根据露点修正的Δ% RH2,且在低湿与高湿时差值较大。
图1. 无线温湿度数据采集仪温度、相对湿度差值图
T发 | % RH发 |
---|---|
20.1 | 19.8 |
20.1 | 29.8 |
20.1 | 39.8 |
20.1 | 49.5 |
20.1 | 59.2 |
20.1 | 68.8 |
20.0 | 78.5 |
20.0 | 88.4 |
表3. 无线温湿度数据采集仪在干湿分离法湿度发生器中测试数据
表3是无线温湿度数据采集仪在干湿分离法湿度发生器中测试数据表,采集仪的温度精度为0.1℃,相对湿度精度为2% RH,性能稳定。
本试验得出的结论如下:
(1) 该修正方法在相对湿度较低时变化较大,20% RH时修正值约为1.4% RH,相对湿度增大后变化趋势减小,90% RH时修正值约为0.8% RH。
(2) 该修正方法无异常值出现,修正前标准器与被校设备的误差在0.6% RH~1.1% RH范围内,修正后误差在0.0% RH~0.4% RH范围内,根据露点修正的相对湿度更接近标准器的值,试验工况下最大误差为0.4% RH;根据标准器示数直接修正的最大误差为1.1% RH。
(3) 未修正和修正相对湿度都在给定的2% RH不确定范围内,因此对温湿度精度要求较高的仪器根据露点修正具有应用价值,对于精度要求不高的仪器可直接通过标准器示数进行修正。
刘宇明,邬 昕. 一种数字式温湿度计相对湿度的修正方法A Correction Method for the Relative Humidity of a Digital Thermohygrometer[J]. 仪器与设备, 2024, 12(01): 68-71. https://doi.org/10.12677/IaE.2024.121010